Measuring / Operation

NitraVis 70x IQ

4 - 8

ba75950e02

11/2012

4.3.3 Sensor check/

Zero adjustment

Note
The sensor check or Zero adjustment must be carried out under abso-
lutely clean conditions. If you do not work carefully enough the Zero ad-
justment can deteriorate the measuring quality.

Ultrapure water

(H2O dist.)

As ultrapure water exclusively use distilled or deionized water that is

suitable for analysis purposes. Especially make sure that the concen-

tration of metrologically relevant substances (nitrate) is negligible.

Storage location of the

Zero adjustment

data

The data of the sensor checks are not stored.

The data of the user zero adjustment are stored in the sensor.

Steps

A successful

Sensor check (H2O dest.)

is the prerequisite for a valid

user calibration. After each

Sensor check (H2O dest.)

the sensor is

evaluated.

A changed sensor evaluation proves that the cleaning was effective.

The steps of the sensor check or

Zero adjustment

are identical except

for the used check solutions (ultrapure water or standard solution) and

are described together below. The sensor check and

Zero adjustment

are carried out with the aid of the sensor sleeve.