Measuring / Operation

NitraVis 70x IQ

4 - 12

ba75950e02

11/2012

Up to this point, you can break off the calibration procedure at any time

with the <ESC> key. The system continues to work with the old calibra-

tion data. The maintenance condition, however, has to be switched off

again in any case.

After starting the calibration with the <OK> key (step 16.) you can no

longer cancel the calibration procedure.

Result

after

Sensor check

(H2O dest.)

If the rating number is not in the zero range:

Repeat the cleaning procedure and the

Sensor check (H2O dest.)

until

the rating number no longer changes significantly.

If the rating number is still outside the zero range, a new zero adjust-

ment is necessary.

Result

after

Zero

adjustment

If the result is erroneous, clean the sensor and all equipment once

again. Then carry out a new

Zero adjustment

and when doing so make

sure the conditions are absolutely clean.

15.

Confirm each checklist with <OK> until the

Start measurement

display appears.

16.

Confirm with <OK>.

The measurement is started. After the end of the measurement

the result appears. With the

Zero adjustment

, the result is

entered in the calibration history at the same time.

Rating number

Meaning

Zero range

(-40 ... +40)

Sensor check (H2O dest.)

successful (OK)

(0 = no deviation)

Positive number

(+40 ... +1000)

Measured values were too low.

Possible cause:

The measurement windows are contaminated.

Negative number

(-40 ... -1000)

Measured values were too high.

Possible cause:

During the last zero adjustment, the measure-

ment windows were contaminated more than dur-

ing the current sensor check.

Result

Meaning

Sensor released for

measurement

OK

Zero adjustment

successful.

Yes

Error

Zero adjustment

not success-

ful.

No ("----")