NitraVis 70x IQ

What to do if...

7 - 3

ba75950e02

11/2012

Measured values too low

or too high

Display of

ERROR

Cause

Remedy

– Gas bubbles in front of the

measurement windows

– Check the installation position

of the sensor

– Measurement windows dirty

– Clean the measurement win-

dows

Cause

Remedy

– UV-VIS sensor not correctly

connected

– Check the assignment of the

terminal connections

– Load at a power supply mod-

ule too high

– Install another power supply

module in the vicinity of the

sensor

– IQ S

ENSOR

N

ET

cable to-

wards the NitraVis 70x IQ too

long (voltage drop too great)

– Install another power supply

module in the vicinity of the

sensor

– Electrical connection be-

tween power supply module

and NitraVis 70x IQ is inter-

rupted

– Check the cable connection

step-by-step starting with the

power supply module, and

replace any defective cable

sections.

– Check the contacts on the MIQ

modules (stack mounting).

Clean any dirty contacts. Care-

fully bend back contact springs

that have been pressed flat or

bent (pay attention to sufficient

spring tension)