3 installation, 1 mounting the sensor, 3 installation -2 – YSI NiCaVis 705 IQ User Manual

Page 12: Mounting the sensor -2

Advertising
background image

Commissioning

NiCaVis 705 IQ

3 - 2

ba75950e01

03/2013

3.3

Installation

3.3.1 Mounting the sensor

Note
The sensor is an optical precision instrument that is to be protected
against shocks. Make sure the distances to the wall, bottom and other
installation surroundings are sufficient. If necessary, protect the sensor
from hitting with the protective ring provided. Mount any fixing elements
only on the stable metal shaft.

Note
Always make sure the surface of the enclosure is not damaged by
sharp-edged objects. The enclosure of the sensor may not touch any
metallic objects in order to prevent contact corrosion.

Note
Do not suspend the sensor on the sensor connection cable. Risk of ca-
ble break and water penetration at the cable gland.

A wide range of mounting accessories is available for the installation of

the NiCaVis 705 IQ (see chapter 6 S

PARE

PARTS

,

MAINTENANCE

EQUIP

-

MENT

,

ACCESSORIES

). Thus, all prevalent ways of mounting are

possible.
Generally, observe the following general rules when mounting the

sensor at the measuring location:

Horizontal installation

When mounting the sensor horizontally, the measuring gap opening

should be on the side. Thus, any air bubbles can escape upward after

cleaning and no sediment can collect in the measuring gap. The sensor

should be aligned in parallel with the flow direction.

Fig. 3-1 Horizontal installation

Measuring gap opening