NiCaVis 705 IQ

Technical data

8 - 1

ba75950e01

03/2013

8

Technical data

8.1

Measurement characteristics

Measuring principle

Spectrophotometric absorption measurement; integrated micropro-

cessor electronics, shielded 2-wire connection for

power and data transmission.

Wavelength range

200 - 720 nm

Measuring gap (optical

layer thickness)

Nitrate measurement,

5 mm gap width

Carbon measurement,

5 mm gap width

The quoted measuring ranges are nominal measuring ranges that are

theoretically possible. In practice, real measuring ranges exist that are

given by the limits of photometric determination. The limits are signifi-

cantly influenced by the light scattering due to solids and the absorption

of accompanying substances (sample matrix). Therefore, no fixed

measuring range can be determined beforehand.

NiCaVis 705 IQ

5 mm

Measuring

location

Measuring

mode

Measuring range

Resolution

Outlet

NO3

0.0 ... 250.0 mg/L

0.1 mg/L

NO3-N

0.00 ... 50.00 mg/L

0.01 mg/L

Measuring

location

Measuring mode

Measuring range

Resolution

Outlet

COD spectr., total

0.0 ... 800.0 mg/L

0.1 mg/L

COD spectr., dis

-

solv

0.0 ... 800.0 mg/L

0.1 mg/L

TOC spectr., correl.

0.0 ... 500.0 mg/L

0.1 mg/L

DOC spectr., correl.

0.0 ... 500.0 mg/L

0.1 mg/L

BOD spectr., correl.

0.0 ... 500.0 mg/L

0.1 mg/L

SAC 254, total

0.0 ... 600.0 1/m

0.1 1/m

SAC 254, dissolved

0.0 ... 600.0 1/m

0.1 1/m

UVT 254

0.0 ... 100.0 %

0.1 %