MIQ/JBR

Installation

3 - 3

ba76077d01

01/2012

Terminator switch

For the setting of the terminator switches within a signal range, the
same rules apply as for systems without MIQ/JBR (see chapter
I

NSTALLATION

of the IQ S

ENSOR

N

ET

system operating manual).

The longest cable section of each signal range is determined.
On both ends of the longest cable section of a signal range the
terminator switch must be set to On. All other terminator switches must
be set to Off.

Fig. 3-3

Settings of the SN terminator switch
for an IQ S

ENSOR

N

ET

system with 2 MIQ/JBR modules

Note
The proceeding to determine the longest cable section can be found in
chapter I

NSTALLATION

of the IQ S

ENSOR

N

ET

system operating manual.

The proceeding to set the terminator switches can also be found in
chapter I

NSTALLATION

of the IQ S

ENSOR

N

ET

system operating manual.

x

x

A

B

x

Signal range 1

Signal range 2

Signal range 3

MIQ/JBR

MIQ/JBR

A

B

x

x

x

= MIQ module with SN terminator switch On

= MIQ module with SN terminator switch Off

= longest cable section of a signal range