5 technical data, Technical data -1, Chapter 5 t – YSI IQ SensorNet MIQ/JBR Module User Manual
Page 15: Echnical, Data, Must, 5technical data

MIQ/JBR
Technical data
5 - 1
ba76077d01
01/2012
5
Technical data
Note
General technical data on MIQ modules are given in the T
ECHNICAL
DATA
chapter of the IQ S
ENSOR
N
ET
system operating manual.
Electrical data
Instrument safety
Terminal connections
Nominal voltage
Max. 24 VDC
via the IQ S
ENSOR
N
ET
(for details, see
chapter T
ECHNICAL
DATA
of the
IQ S
ENSOR
N
ET
system operating manual).
Power consumption
approx. 0.2 W
Protective class
III
Number of MIQ/JBR
modules in an
IQ S
ENSOR
N
ET
system
2
Applicable norms
– EN 61010-1
– UL 3111-1
– CAN/CSA C22.2 No. 1010.1
IQ S
ENSOR
N
ET
connections
4
2 additional switchable SENSORNET
terminators (terminating resistors)
Terminal type
Screw-type terminal strip, accessible by
raising the lid
Terminal ranges
Solid wires:
Flexible wires:
0.2 ... 4.0 mm
2
AWG 24 ... 12
0.2 ... 2.5 mm
2
Cable feeds
4 cable glands M16 x 1.5 on the underside
of the module