8 indexes, 1 explanation of the messages, 1 error messages – YSI IQ SensorNet MIQ/R6 Module User Manual

Page 53: 2 info messages, Indexes -1, Explanation of the messages -1 8.1.1, Error messages -1, Info messages -1, 8indexes

Advertising
background image

MIQ/R6

Indexes

8 - 1

ba76031e01

01/2012

8

Indexes

8.1

Explanation of the messages

In this chapter you will find a list with all the message codes and
corresponding message texts that may occur in the log book of the
IQ S

ENSOR

N

ET

system for the MIQ/R6 output module.

Note
Information about

Contents and structure of the log book and

Structure of the message code

can be found in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system

operating manual.

Note
All message codes of the MIQ/R6 output module end with the number
"431".

8.1.1

Error messages

8.1.2

Info messages

The MIQ/R6 output module does not send any info messages.

Message code

Message text

EI2431

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal connections and module connections
* Defective components,
replace components

Advertising