2 system monitoring, See section 4.5.2 – YSI IQ SensorNet MIQ/R6 Module User Manual

Page 30

Advertising
background image

Settings

MIQ/R6

4 - 16

ba76031e01

01/2012

4.5.2

System monitoring

Function

The System monitoring enables to monitor system errors.

In order to set up the System monitoring function for a relay output, the
relay output must not be linked with a sensor (see section 4.3).

It can be used to monitor the following system errors.

Settings

Collective error

message

In the case of a Collective error message, the relay opens if one of the
following malfunctions occurs:

One of the sensors properly registered at the controller does not
supply a valid main measured value

One of the sensors properly registered at the controller does not
supply a valid secondary measured value

The monitoring MIQ/R6 has not received any new data from the
controller for two minutes.

In any case, the relay remains open for 10 seconds and only closes
when the malfunction is no longer present.

Settings

Selection

Explanations

Power failure

On
Off

The Power failure On function
monitors the supply voltage of
the IQ S

ENSOR

N

ET

at the MIQ/

R6. If the voltage falls below the
critical value, the relay switches.

Communication

On
Off

The Communication On function
monitors the functioning of the
controller and the
communication to the MIQ/R6.

Collective error
message

On
Off

The Collective error message
function simultaneously
monitors the proper functioning
of all sensors and the
functioning of the monitoring
MIQ/R6. (for details see below
this table)

Action

Open

For all functions of the System
monitoring
, the relay action is
permanently set to Open.