1 switching on the maintenance condition, 2 switching off the maintenance condition, Switching on the maintenance condition . 4-32 – YSI IQ SensorNet MIQ/R6 Module User Manual

Page 46: Switching off the maintenance condition . 4-32, Ee section 4.8.2, Section 4.8.1), See section 4.8.2

Advertising
background image

Settings

MIQ/R6

4 - 32

ba76031e01

01/2012

4.8.1

Switching on the maintenance condition

4.8.2

Switching off the maintenance condition

1

Call up the measured value display with

m

.

2

Select the sensor you want to switch on the maintenance
condition for with

d

. The measured values of the sensor do

not flash.

3

Open the Display/Options menu with

g

.

4

Highlight the Switch maintenance condition on/off menu item
with

d

and confirm with

g

. A window that informs you about

the maintenance condition opens up.

5

Confirm Continue with

g

.

The selected sensor is in the maintenance condition. Linked
outputs are frozen.

6

Call up the measured value display with

m

.

The measured values of the sensor flash.

1

Call up the measured value display with

m

.

2

Select the sensor you want to switch off the maintenance
condition for with

d

. The measured values of the sensor

flash.

3

Open the Display/Options menu with

g

.

4

Highlight the Switch maintenance condition on/off menu item
with

d

and confirm with

g

. A window that informs you about

the maintenance condition opens up.

5

Confirm Continue with

g

.

The maintenance condition of the selected sensor is switched
off. Linked outputs are released.

6

Call up the measured value display with

m

.

The measured values of the sensor do not flash.