Commissioning

VisoTurb

®

700 IQ (SW)

3 - 6

ba76039e01

01/2012

3.3.2

Measuring in an open channel (range > 100 FNU)

Example:

Outlet of a waste water

treatment plant (open

channel, wall material:

concrete)

In an open channel, the sensor can be immersed in the sample using
a wall mounting assembly, e.g. EH/W 170 wall mounting assembly,
(please note the minimum immersion depth).

 Protect the measuring location and the environment against direct

sunlight (sun shield or similar)

 Mount the sensor rigidly in the channel. At the same time, tilt the

sensor approx. 20 to 45 ° against the direction of the flow.

 Install the sensor so that the marking on the sensor points towards

the outlet of the channel.

Fig. 3-4

Turbidity sensor in the open channel with EH/W 170 fixture assembly for
direct wall mounting.

Note
Interferences at the measuring location (see section 3.2.1) may require
special adaptations of the installation. For exceptions to the direction of
flow, see section 3.2.2 F

LOW

DIRECTION

.

min. 10 cm

min. 10 cm

Ground

Marking
in this
direction

Direction of flo

w

min. 10 cm

Immersion depth min. 7 cm

45 °

Advertising