What to do if ...

VisoTurb

®

700 IQ (SW)

6 - 2

ba76039e01

01/2012

Measured values too

high

Measured value flashes

Cause

Remedy

– Gas bubbles in the medium are

in front of the sapphire disc

– Check the installation position

of the sensor (see section 3.2
and section 3.3)

– Light scattering on the walls

– Check the installation position

of the sensor (see section 3.2
and section 3.3)

– If necessary, compensate for

effects that cannot be
removed with the aid of the
application offset function

Cause

Remedy

– Maintenance condition is

switched on

– if the maintenance condition

was switched on manually
(e.g. by pressing the

c

key):

Switch off the maintenance
condition manually in the
Display/Options menu (see
system operating manual)

– if the maintenance condition

was switched on
automatically (e.g. by the
cleaning system):
The maintenance condition
will be terminated
automatically