2 application characteristics, Application characteristics -2 – YSI VisoTurb 700 IQ User Manual

Page 40

Advertising
background image

Technical data

VisoTurb

®

700 IQ (SW)

7 - 2

ba76039e01

01/2012

Accuracy

7.2

Application characteristics

Allowed

temperature range

Allowed pH range of the

measuring medium

4 ... 12

Pressure resistance

Sensor with connected SACIQ (SW) sensor connection cable:

The sensor fulfills all requirements according to article 3(3) of the
directive 97/23/EC ("pressure equipment directive").

Type of protection

Sensor with connected SACIQ (SW) sensor connection cable:

IP 68, 10 bar (10

6

Pa)

Depth of immersion

min. 10 cm; max. 100 m depth

Operating position

see section 3.2 I

NSTALLATION

Fields of application

Water and wastewater monitoring

Process variation coefficient
according to DIN 38402 part 51

< 1 % in the range to 2000 FNU

Repeatability limit or repeatability
according to DIN ISO 5725 or
DIN 1319 respectively

< 0.015 % or
min. 0.006 FNU.

Measuring medium

0 °C ... + 60 °C (32 ... 140 °F)
Operation with ultrasound
cleaning system possible up to
40 °C

Note: The ultrasound cleaning
system automatically switches
off and on again. The switching
off above 40 °C prevents an
overheating, for example if the
minimum immersion depth of the
sensor is not maintained.

Storage/transport

- 5 °C ... + 65 °C (23 ... 149 °F)

Max. allowed overpressure

10

6

Pa (10 bar)

Max. allowed underpressure

temporarily 5 x 10

4

Pa (0,5 bar)