Commissioning

3 - 11

ba76039e01

01/2012

3.4.2

Setting table

VisoTurb

®

700 IQ (SW)

Setting

Selection/values

Explanation

Measuring mode

FNU

NTU

TEF
mg/l SiO2
ppm SiO2
g/l TSS

– Turbidity unit Formazine Nephelometric

Units

– Turbidity unit Nephelometric Turbidity

Units

Turbidity Unit Formazine

– Concentration of SiO

2

in mg/l

– Concentration of SiO

2

in ppm

– Conc. of total suspended solids in g/l

(for details, see section 4.2.3).

Measuring range

AutoRange
0 ... 0.400 FNU
0 ... 4.00 FNU
0 ... 40.0 FNU
0 ... 400 FNU
0 ... 4000 FNU

Measuring ranges for the FNU measuring
mode

(AutoRange = automatic changeover of the
measuring range)

AutoRange
0 ... 0.400 NTU
0 ... 4.00 NTU
0 ... 40.0 NTU
0 ... 400 NTU
0 ... 4000 NTU

Measuring range for the measuring mode
NTU

(AutoRange = automatic changeover of the
measuring range)

AutoRange
0 ... 0.400 TEF
0 ... 4.00 TEF
0 ... 40.0 TEF
0 ... 400 TEF
0 ... 4000 TEF

Measuring range for the measuring mode
TEF

(AutoRange = automatic changeover of the
measuring range)

AutoRange
0 ... 0.400 mg/l
0 ... 4.00 mg/l
0 ... 40.0 mg/l
0 ... 400 mg/l
0 ... 4000 mg/l

Measuring range for the measuring mode
mg/l SiO2

(AutoRange = automatic changeover of the
measuring range)