9.1

Explanation of the messages

This chapter contains a list of all the message codes and related
message texts that can occur in the log book of the IQ S

ENSOR

N

ET

system for the VisoTurb

®

700 IQ (SW) sensor.

Note
Information on

 the contents and structure of the log book and

 the structure of the message code

is given in the L

OG

BOOK

chapter of the IQ S

ENSOR

N

ET

system

operating manual.

Note
All message codes of the VisoTurb

®

700 IQ (SW) end with the number

"341".

9.1.1

Error messages

Message code

Message text

EA2341

Sensor temperature too high!
* Check process and application

EA3341

Sensor temperature too low!
* Check process and application

EA6341

Meas. range exceeded or undercut
* Check process
* Select other meas. range
* Submerse sensor in sample
* Select bubble-free spot for measurement
* Remove any foreign matter from sensor
* Avoid influence of large foreign matter
* Clean sensor
* Increase signal average time

EA7341

Ultrasound cleaning system switched off
* Check sample temperature
* Submerse sensor in sample

EI1341

Operational voltage too low
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Defective components, replace components