2 info messages, Info messages -2 – YSI VisoTurb 700 IQ User Manual

Page 46

Advertising
background image

Indexes

VisoTurb

®

700 IQ (SW)

9 - 2

ba76039e01

01/2012

9.1.2

Info messages

EI2341

Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components

ES1341

Component hardware defective
* Contact service

ESD341

SensCheck: Measurement interfered
* Submerse sensor in sample
* Select bubble-free spot for measurement
* Remove any foreign matter from sensor
* Avoid influence of large foreign matter
* Clean sensor
* Increase signal average time

ESE341

SensCheck: Ultrasound cleaning system has failed
* Return sensor for repair

Message code

Message text

Message code

Message text

IA1341

Ultrasound cleaning system switched on
* Check sensor visually
* Clean sensor if necessary