Info messages
EI2341
Operational voltage too low, no operation possible
* Check installation and cable lengths, Follow installation instructions
* Power unit(s) overloaded, add power unit(s)
* Check terminal and module connections
* Defective components, replace components
ES1341
Component hardware defective
* Contact service
ESD341
SensCheck: Measurement interfered
* Submerse sensor in sample
* Select bubble-free spot for measurement
* Remove any foreign matter from sensor
* Avoid influence of large foreign matter
* Clean sensor
* Increase signal average time
ESE341
SensCheck: Ultrasound cleaning system has failed
* Return sensor for repair
Message code
Message text
Message code
Message text
IA1341
Ultrasound cleaning system switched on
* Check sensor visually
* Clean sensor if necessary