Sensor storage, Troubleshooting – YSI Pro30 User Manual

Page 14

Advertising
background image

20

21

sensor storage
SHoRT AND LoNG TERM SToRAGE

For both short and long term storage, the conductivity sensor should be stored

clean and dry.

Remove the batteries from the instrument when storing it for long periods of

time (>30 days).

Long Term Storage Temperature: -5 to 70°C (23 to 158°F)

T R o U B L E S H o o T I N G

Symptom

Possible Solution

Instrument will not turn on,

a battery symbol appears,

or “Critical Shutdown”

displays on the screen.

Low battery voltage, replace batteries.

1.

Batteries installed incorrectly,

2.

check battery polarity.

Return system for service.

3.

Temperature values display

over or Undr on Run screen.

Sample temperature is less than -5° C or

1.

more than +55°C. Increase or decrease

the sample temperature to bring within

the allowable range.

Contact YSI Tech Support.

2.

Instrument will not calibrate the

Conductivity sensor; instrument

displays “Calibration over”,

“Calibration Under”, or “Unstable

Reading” during calibration.

Ensure the conductivity sensor is

1.

clean. Follow the cleaning procedures

in the Care, Maintenance and

Storage section of this manual.

Verify the calibration solution is above

2.

the two holes near the cable, see figure 8.

Verify the calibration solution

3.

is not expired or contaminated.

Try a new bottle of solution.

Ensure you are entering in the correct

4.

value for the solution according to the

measurement units. 1 mS = 1,000 uS.

Allow sufficient stabilization time

5.

for conductivity and temperature

AND wait at least 3 seconds before

confirming a calibration.

Contact YSI Tech Support.

6.

Symptom

Possible Solution

Conductivity readings

are inaccurate.

Ensure the conductivity sensor is

1.

clean. Follow the cleaning procedures

in the Care, Maintenance and

Storage section of this manual.

Verify the sample is above the two

2.

holes near the cable, see figure 8.

Verify calibration.

3.

Verify temperature readings

4.

are accurate.

Verify the correct units are setup in the

5.

System Setup menu, i.e. uS vs mS and

Conductivity vs. Specific Conductance.

Contact YSI Tech Support.

6.

Conductivity values display

over or Undr on Run screen.

Ensure the conductivity sensor is

1.

clean. Follow the cleaning procedures

in the Care, Maintenance and

Storage section of this manual.

Verify the sample is above the two

2.

holes near the cable, see figure 8

Verify calibration.

3.

Verify temperature readings

4.

are accurate.

Sample conductivity is outside

5.

the measurement range of the

instrument, i.e. 0-200 mS.

Contact YSI Tech Support.