Conductivity units (cond. units), Conductivity auto stable (cond. auto stable), Temperature units – YSI Pro30 User Manual

Page 8

Advertising
background image

8

9

conductIvIty unIts (cond. unIts)

Highlight Cond. Units (Conductivity Units) and press enter to open a submenu

that allows you to select the conductivity units to be displayed on the Run screen.

Highlight a unit and press enter to enable or disable it. An enabled conductivity

unit will have an ‘X’ in the box next to it. Highlight the ESC-Exit box along

the bottom of the display and press enter to save any changes and to close the

conductivity units submenu.

There are seven options for displaying conductivity. only four units can be

enabled at the same time:

CoND-mS/cm displays conductivity in milliSiemens per centimeter.

CoND-uS/cm displays conductivity in microSiemens per centimeter.

SPC-mS/cm displays Specific Conductance in milliSiemens per

centimeter. Specific Conductance is temperature compensated

conductivity.

SPC-uS/cm displays Specific Conductance in microSiemens per

centimeter. Specific Conductance is temperature compensated

conductivity.

Sal ppt displays salinity in parts per thousand. The salinity reading is

calculated from the instrument’s conductivity and temperature values

using algorithms found in Standard Methods for the Examination of

Water and Wastewater.

TDS g/L displays

Total Dissolved Solids in grams per liter. TDS is

calculated from conductivity and temperature using a user-selectable

TDS constant.

TDS mg/L displays

Total Dissolved Solids in milligrams per liter. TDS

is calculated from conductivity and temperature using a user-selectable

TDS constant.

Note: 1 milliSiemen = 1,000 microSiemens.

SPECIFIC CoNDUCTANCE

The conductivity of a sample is highly dependent on temperature, varying as

much as 3% for each change of one degree Celsius (temperature coefficient =

3%/°C). In addition, the temperature coefficient itself varies with the nature of

the ionic species present in the sample. Therefore, it is useful to compensate for

this temperature dependence in order to quickly compare conductivity readings

taken at different temperatures.

The Pro30 can display non-temperature compensated conductivity as well as

temperature compensated Specific Conductance. If Specific Conductance is

selected, the Pro30 uses the temperature and conductivity values associated with

each measurement to calculate a specific conductance value compensated to a

user selected reference temperature, see below. Additionally, the user can select

the temperature coefficient from 0% to 4%.

Using the Pro30’s default reference temperature and temperature coefficient (25

°C and 1.91%), the calculation is carried out as follows:

Specific Conductance (25°C) = Conductivity of sample

1 + 0.0191 * (T - 25)

T = Temperature of the sample in °C

conductIvIty auto staBle (cond. auto staBle)

Auto Stable utilizes preset values to indicate when a reading is stable. The preset

values are adjustable in the System Setup menu. The user can input a % change

in readings (0.0 to 1.9) over ‘x’ amount of time in seconds (3-19).

Highlight Cond. Auto Stable, then press enter to open the submenu.

Use the up or down arrow keys to highlight the % change or seconds (secs) input

field, then press enter to make the highlighted field adjustable. Use the up or

down arrow keys to adjust the selected value, then press enter to confirm changes.

once you have confirmed any changes, highlight the ESC-Exit box along the

bottom of the display and press enter to close the Auto Stable submenu.

To disable Auto Stable, set the % Change input to 0.0.

When Auto Stable is enabled, an

AS

symbol will display next to the reading

on the Run screen and blink during stabilization. When the dissolved oxygen

and/or conductivity reading stabilizes based on the Auto Stable settings, the

AS

symbol will display steadily and the instrument will beep twice if Audio

is turned on.

temPerature unIts

Highlight Temperature Units and press enter to open a submenu that allows

you to change the temperature units displayed on the Run screen. Highlight

the desired unit (Celsius or Fahrenheit) and press enter to enable. The enabled

temperature unit will have an ‘X’ in the box next to it. only one unit may be

enabled at a time. Highlight the ESC-Exit box and press enter to save any changes

and to close the Temperature Units submenu.

Advertising