H-3521 – Xylem H-3521 User Manual
Page 40

5-10 SDI-12 Command and Response Protocol
H-3521
5.9 Extended Set Current Stage Command
The H-3521 processes the pressure sensor input and computes Pressure in PSI units. Stage is
computed with a Stage = m * Pressure + b equation. During installation it is convenient to
quickly set the H-3521's Stage reading to match the current stage, elevation or pressure of the
water as determined by a staff gauge or other datum. This command causes the H-3521 to make a
fresh measurement and automatically update the Offset (b) term as needed to produce the desired
Stage
.
Command
Response
Description
"aXSCSddd!"
“a0061<cr><lf>"
Set current Stage
Where: a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
XSCS
are upper case characters.
ddd
is the new current stage value.
Example of a H-3521 Extended "Set Current Stage" command:
Command
Response
Time Values
Description
"aXSCS2.3!"
"a0041<cr><lf>"
4 sec 1
Set the Stage to 2.3
Subsequent Command
Response
Description
"aD0"
a+1.80<cr><lf>
The new Offset