H-3521 – Xylem H-3521 User Manual

Page 45

Advertising
background image

H-3521

SDI-12 Command and Response Protocol 5-15

5.14 Extended Read Measure Rate and Write Measure Rate
When the H-3521 is put in power mode 3 it makes continuous measurements based on a measure
rate. This command is for reading/writing that measure rate value. The measure rate range is 0-
255 minutes. When the user sets the measure rate to 0 then the H-3521 makes continuous
measurements as fast as the H-3521 will measure, which is about 6 seconds by default. The
factory default is 1 minute measure rate.

Command

Response

Description

"aXRMR!"

“a0011<cr><lf>"

Read Measure Rate

"aXWMRnnn!"

“a0021<cr><lf>"

Write Measure Rate

Where:

a

is the sensor address ("0-9", "A-Z", "a-z", "*", "?").

XRMR

are upper case characters.

XWMR

are upper case characters.

nnn

is the rate in minutes that the sensor will update all outputs.

Example of a H-3521 Extended "Read Measure Rate" command:
Command

Response

Time

Values

Description

"aXRMR!"

"a0011<cr><lf>"

1 sec

1

Read Measure Rate

Command

Response

Description

"aD0!"

"a+1<cr><lf>"

Measure Rate = 1 minute

Example of a H-3521 Extended "Write Measure Rate" command:
Command

Response

Time

Values

Description

"aXWMR2!"

"a0021<cr><lf>"

2 sec

1

Write Measure Rate

Command

Response

Description

"aD0!"

"a+2<cr><lf>"

Measure Rate = 2 minutes