H-3521 – Xylem H-3521 User Manual
Page 42

5-12 SDI-12 Command and Response Protocol
H-3521
5.11 Extended Read/Write Stage Digits
This is for reading or writing the stage digits to the right of the decimal point. The stage digits by
default is set to 2.
Command
Response
Description
“aXRSD!”
“a0011<cr><lf>”
Read Stage Digits
“aXWSDd!” “a0021<cr><lf>”
Write Stage Digits
Where:
a
is the sensor address ("0-9", "A-Z", "a-z", "*", "?").
XRSD
are upper case characters.
XWSD
are upper case characters.
d
is the new enable value(0-9)
Example of a H-3531 Extended "Read Stage Digits" command:
Command
Response
Time Values
Description
"aXRSD!"
"a0011<cr><lf>"
1 sec
1
Read Stage Digits
Command
Response
Description
"aD0!"
"a+2<cr><lf>"
Stage digits = 2
Example of a H-3531 Extended "Write Stage Digits" command:
Command
Response
Time Values
Description
"aXWSD3!"
"a0021<cr><lf>"
2 sec
1
Set Stage digits to 3
Command
Response
Description
"aD0!"
"a+3<cr><lf>"
Stage digits = 3