4.4.3 Logging Mode Option

Logging Mode, Coefficients, and User Notes configure the “

W

ATER

LOG” for an unattended

logging session. The Logging Mode Option is shown in Figure 4-8. The logging rate indicates the
elapsed time between measurements. The “

W

ATER

LOG” clock circuitry allows the

W

ATER

LOG” to enter a low-power “SLEEP” mode of operation between measurements. When

the time interval has elapsed, which is specified in Logging Mode, the “

W

ATER

LOG” will

“WAKE-UP” and take a pressure, temperature and battery voltage measurement. It will then
store the data in FLASH EEPROM with a time-stamp, and then go to “SLEEP” again.

Figure 4-8. Logging Mode Option

The time between samples will range from 1 second up to 24 hours. If the logging rate is less
than 6 seconds, the “

W

ATER

LOG” when running, will not go to “SLEEP”.

The “

W

ATER

LOG” “WAKES-UP” and takes a measurement when the time in the DATE/TIME

field matches the “Next Log Time”. The “Next Log Time” is incremented by the logging rate. If