Dh-21 – Xylem DH-21 User Manual
Page 68

4-22 Using the Graphical “W
ATER
LOG” Software
DH-21
SCAN RATE
HYPER DIVISOR
FAST SCAN RATE
1 minute
60
1 seconds
1 minute
30
2 seconds
1 minute
20
3 seconds
1 minute
15
4 seconds
1 minute
12
5 seconds
1 minute
10
6 seconds
1 minute
6
10 seconds
1 minute
5
12 seconds
1 minute
4
15 seconds
1 minute
3
20 seconds
1 minute
2
30 seconds
1 minute
1
60 seconds
This option allows the sensor to capture and record quick changing data without the expense of
battery life by adjusting the scan rate based on the data change rate. It also saves memory by not
logging data when the data is not changing significantly.
Note that this check is on the level parameter only, not the temperature, and the values entered
for the hyper limit and delta limit will be in the same units as set in the coefficients options. For
example, if the coefficients Pressure / Level option is set to feet, the hyper limit and delta limit
will also be in feet.