Required software

The EBI16 is supported:
-

Winlog.med

V 3.5

-

Winlog.validation

V 3.5

Logger and interface

-

To program or to start recording the
EBI16 must be placed on the
interface.

-

To read out the data of the EBI16 it
must be placed on the interface.

-

Make sure that the case of the EBI16
is not hotter than +70°C. It could
damage the housing of the interface
otherwise

Cool down times

-

To prevent damage to the housing of
the interface, the EBI16 must cool
down for approx. 7 minutes before it
can be placed on the interface.

-

To reach the start condition of
+15°C … +35°C -for programming-
after a cycle with 134°C the EBI16
must cool down at room temperature
for approx. 45 to 60 minutes.

Logiciel requis

L’EBI 16 peut être utilisé avec :
-

Winlog.med

V3.5

-

Winlog.validation

V3.5

Enregistreur et interface

-

Pour

programmer

ou

démarrer

l’enregistrement, l’EBI 16 doit être
placé sur l’interface

-

Pour lire les données de l’EBI 16, il
faut

également

le

placer

sur

l’interface

-

Attention à ce que la température du
boîtier de

l’EBI 16 n’excède pas

+70°C. Cela pourrait endommager le
boîtier le l‘interface

Durée de refroidissement

-

Afin d’éviter tout dommage au niveau
du boîtier de l’interface, l’EBI 16 doit
refroidir environ 7 minutes avant
d’être positionnés sur l’interface.

-

Lo

rs d’un cycle à +134°C, pour

atteindre la condition de départ entre
15°C et 35°C, il faut laisser reposer
l’EBI 16 environ 45mn à 60mn à
température ambiante.

Advertising