Do module: cleaning the electrodes – Xylem LFOx1400_LFOx1400-ID User Manual

Page 15

Advertising
background image

LFOx1400(-ID)

Maintenance, cleaning, replacement

39

ba75523e04

07/2011

5.4

DO module: Cleaning the electrodes

Caution

Before starting to work with the sensor, please note the G

ENERAL

MAINTENANCE

INSTRUCTIONS

on page 34.

General

information

Cleaning is only required in cases of slopes too small or too large (sensor
cannot be calibrated) that cannot be resolved by changing the membrane cap
and electrolyte solution.

Cleaning the electrodes

Unscrew the membrane cap.

Caution: Electrolyte solution!
For disposal of the membrane
cap and electrolyte solution,
see section 5.7.

Rinse the sensor head with
deionized water.

Using the rough side of the
wet SF300 polishing strip,
polish off any contamination
from the gold working elec-
trode using light pressure.

Caution: Do not use any con-
ventional sandpaper or glass-
fiber brushes.

Rinse the sensor head with
deionized water.

Wipe the counter electrode
with a lint-free paper towel
and carefully remove any
loose white deposits.

Immerse the sensor head in-
cluding the counter electrode
in Ox921 cleaning solution.

Allow to react for 1 to 3 min-
utes.

SF300

Ox921