LFOx1400(-ID)

Commissioning

31

ba75523e04

07/2011

3.1.2

Scope of delivery of the conductivity basic module

 Conductivity basic module
 1 closing cap, mounted on the conductivity basic module
 Operating manual for LFOx1400(-ID)

3.1.3

Scope of delivery of the DO module

 DO module, filled with electrolyte solution and operable
 1 closing cap, mounted on the DO module
 1 exchange membrane cap, Ox923
 Polishing strip, SF300
 Electrolyte solution, Ox920
 Cleaning solution, Ox921
 Operating manual for LFOx1400(-ID)

Note
The membrane cap that is mounted on the DO sensor for delivery serves
mainly as a transport protection. Depending on the duration of the transport
and storage period, it may have a shortened operational lifetime. If the mea-
suring system cannot be calibrated (error message on the instrument),
please proceed according to section section 5.3 DO

MODULE

: C

HANGING

THE

ELECTROLYTE

SOLUTION

AND

MEMBRANE

CAP

.