Technical data

LFOx1400(-ID)

46

ba75523e04

07/2011

Materials

Working electrode

Gold

Counter electrode

Lead

– Membrane cap
– Cap nut

POM

Membrane

FEP, 13 µm

– Sensor head

Epoxy, PEEK

– Plug connection

PEEK

Screw thread base

Stainless steel 1.4571

Seals

FPM (Viton)

Closing cap

POM

Plug connection

Plug type

3-pole, watertight plug connection with cap nut,
reverse polarity protected, contacts gold-plated

Pressure

resistance

Sensor with closed plug
connection

IP 68 (3 x 10

5

Pa or 3 bar)

The DO module meets the requirements according to article 3(3) of the 97/
23/EC directive ("Pressure equipment directive").

Measurement

conditions

Measuring ranges at
20 °C

0 ... 50 mg/l DO
0 ... 600 % DO saturation
0 ... 1250 mbar DO partial pressure

Polarization time

min. 30 ... 50 min after changing the electrolyte,
it is not necessary to connect the sensor to the
measuring instrument for polarization

Temperature range

0 ... 50 °C

Max. allowed overpres-
sure

3 x 10

5

Pa (3 bar)

Depth of immersion with
conductivity basic module

min. 8 cm
max. 20 m (depending on the cable length of
the conductivity basic module)

Operating position

any

Approach flow

> 3 cm/s at 10 % measurement accuracy
10 cm/s at 5 % measurement accuracy
18 cm/s at 1 % measurement accuracy

Storage

conditions

Recommended storing
method

in the OxiCal

®

-CX calibration vessel (moist-

ened)

Storage temperature

0 ... 50 °C