Xylem ConOx User Manual
Page 13
 
ConOx
Maintenance, cleaning, replacement
37
Carefully shake off the drops 
of water.
Fill a new membrane cap with 
ELY/G electrolyte solution.
Remove any air bubbles by 
carefully tapping the mem-
brane cap. Additionally, you 
can prevent air bubbles by 
throwing the first filling away 
and refilling the membrane 
cap.
Thoroughly rinse the sensor 
head with electrolyte solution.
Hold the sensor inclined and 
screw on the membrane cap 
fingertight using a paper tow-
el. Excess electrolyte solution 
is forced out of the ventilation 
area.
Check the filling:
Inspect the face surface. No 
air bubbles may be present 
within the dashed circle. Air 
bubbles outside this area do 
not interfere.
Note
For measurements under high pressure the filling must be com-
pletely free of air bubbles.
Readiness to measure
The DO module is ready for operation after approx. 30 to 50 min-
utes. Subsequently calibrate the sensor for DO measurements.
Note
If you want to measure very low DO concentrations (< 0.5 % satu-
ration), we recommend to let the sensor rest overnight and then 
calibrate it.
EL
Y/G
EL
Y/G
Ventilation area
points upwards
4 mm