2 dissolved oxygen (do) module, Dissolved oxygen (do) module – Xylem ConOx User Manual

Page 20

Advertising
background image

Technical data

ConOx

44

7.2

Dissolved oxygen (DO) module

Pressure

resistance

Sensor with closed plug
connection

IP 68 (3 x 10

5

Pa or 3 bar)

Cable plug

IP 67 (when plugged in)

The ConOx meets the requirements according to article 3(3) of the
97/23/EC directive ("Pressure equipment directive").

Measurement

conditions

Conductivity measuring
range

1

m

S/cm ... 2 S/cm

Temperature range

0 ... 50 °C

Max. allowed overpressure

3 x 10

5

Pa (3 bar)

Depth of immersion

min. 6 cm (with closing cap)
min. 8 cm (with DO module)
max. 20 m (depending on the cable length)

Operating position

any

Approach flow

not required

Storage

conditions

Recommended storing
method

in the OxiCal

®

-CX calibration vessel (moist-

ened)
or dry (without DO module)

Storage temperature

0 ... 50 °C

Characteristic data

on delivery

Temperature responding
behavior

t

99

(99 % of the final value display after)

< 20 s

Precision of the temperature
sensor

± 0.3 K

General features

Measuring principle

Membrane covered galvanic sensor

Temperature compensa-
tion

IMT compensation
(calculated by the measuring instrument)

Dimensions

(in mm)

Weight

approx. 20 g (filled with electrolyte)