4 measuring / operation, 1 calibration, 2 measuring – Xylem ConOx User Manual

Page 9: Measuring / operation, Calibration, Measuring, 4measuring / operation

Advertising
background image

ConOx

Measuring / Operation

33

Unscrewing the

DO module

Unscrew the cap nut and remove the DO module from the conductivity basic
module.

Caution

Inappropriate handling can lead to the release of electrolyte solution.
When unscrewing the DO module only turn the cap nut (not the
membrane cap!).

3.2.3

Preparing the measuring operation

Connect the sensor to the measuring instrument. The sensor is immediately
ready to measure. It is not necessary to polarize the DO module.

4

Measuring / Operation

4.1

Calibration

To calibrate for DO measurements, use the OxiCal

®

-CX calibration and

storage vessel. If the A 325/K or A 325/S armoring is used, you have to
remove the protective hood and fixing nut of the armoring first.

When determining the cell constant, observe the respective minimum
immersion depth with the closing cap or D. O. module screwed on (important:
the temperature sensor must be submersed). The minimum immersion
depths are given in chapter 7 T

ECHNICAL

DATA

.

Note
Please read the course of the calibration in the operating manual of the mea-
suring instrument.

4.2

Measuring

Please always observe the required minimum immersion depth and the
minimum approach flow that is important for DO measurements (see
chapter 7 T

ECHNICAL

DATA

).

Minimum

approach flow for

DO measurements

The minimum flow can be provided in different ways, e. g.:

l

The flow of the water to be measured is sufficient (aeration tank, water
pipe, stream)

l

Slowly pull the sensor through the water by hand (lake, container), or

l

Use a flow aid, e. g. a magnetic stirrer with stirring device or a battery stir-
rer (see chapter 8 W

EAR

PARTS

AND

ACCESSORIES

)

Note
When using the BR 325 battery stirrer, please make sure that the protective
pipe already has the groove in the area of the conductivity basic module (from
the year of manufacture, 2004).