Maintenance, cleaning, replacement

FDO

®

925

28

ba75898s01

04/2010

Note
If the sensor cap is visibly damaged it has to be replaced.

3.4

Checking the zero point of the sensor

There are two methods to check the zero point of the sensor:

z Measurement in a nitrogen atmosphere (recommended method)
z Measurement in a sodium sulfite solution according to DIN EN 25814/

ISO 5814.

Test criterion

The sensor is OK if the measuring instrument displays < 0.5 % D.O.
saturation after 15 minutes.