Xylem FDO 925 User Manual

Page 15

Advertising
background image

FDO

®

925

Technical data

31

ba75898s01

04/2010

Measurement

conditions

Measuring ranges at 20 °C
(68 °F)

0 ... 20 mg/l D.O.
0 ... 200 % D.O

.

saturation

0 ... 400 mbar D.O. partial pressure

Temperature range

0 ... 50 °C (32 ... 122 °F)

Max. admissible overpres-
sure

2.5 x 10

5

Pa (2.5 bar)

Immersion depth

min. 6 cm
max. 25 m (depending on the cable length)

Operating position

any

Approach flow

not required

Storage conditions

Recommended storing
method

in the check and storage vessel,
FDO

®

Check

Storage temperature

0 ... 50 °C (32 ... 122 °F)

Characteristic data

on delivery

Zero signal

< 0.1 % of the saturation value

Response time at 20 °C
(68 °F) in stirred solution

t

90

(90% of the final value display after) < 30s

t

95

(95% of the final value display after) < 45s

t

99

(99% of the final value display after) < 60s

Response time of tempera-
ture measurement

t

99

(99% of the final value display after) < 60s

Precision of temperature
measurement

± 0.2 K

Working life of the sensor
cap

Min. 1 year with authorized use