Measuring / Operation

FDO

®

925

24

ba75898s01

04/2010

2.4

Function check and user calibration

Factory calibration

The FDO

®

925 is factory calibrated. In the recommended application (see

page 22), the measuring characteristics of the sensor cap remain stable for
the specified service life. Thus, a user calibration is not usually required.

When does a func-

tion check or user

calibration make

sense?

A function check or user calibration can be useful in the following special
cases:

z If the measured values appear to be implausible and it is assumed that the

service life of the sensor cap is over

z Routinely within the framework of the company quality assurance.

Check or calibra-

tion medium

The check and user calibration take place in water vapor-saturated air. The
suitable conditions are easily provided with the aid of the check and calibra-
tion vessel, FDO

®

Check. For this purpose, moisten the sponge inside the

vessel. Then insert the sensor in the vessel as far as it will go. The sensor
membrane must be clean and dry for this.

Moisten the sponge:

z Remove the cap (1).
z Take out the sponge (2), wet it, then slightly squeeze it out.
z Insert the sponge again and close the calibration and storage vessel with

the lid.

After inserting the sensor, wait for the temperatures of the sensor and
calibration vessel to adjust.

Note
The steps of the check or user calibration are described in detail in the oper-
ating manual of the meter.

2.5

Storage

Always store the sensor in the check and storage vessel (FDO

®

Check) at a

temperature in the range 0 ... 50 °C (32 ... 122 °F).