Wear parts and accessories

TruLine

Oxidizing
pretreatment

Reducing
pretreatment

Immerse the platinum electrode for two to three

days in a sulfuric acid chlorine solution. Chlorine
powder f or producing the solution is included in the
SORT/RH reagent set.

Note: The diaphragm must not be immersed in the
chlorine solution!

W hen the electrode is ready f or the test immerse it
in the Zobell ORP buffer solution and wait f or a sta-
ble measured value.

Wear parts and accessories

Description

Model

Order no.

Storage electrolyte solution 3 mol/l KCl,
Ag

+

-free (250 ml)

KCl-250

109 705Y

Reagent set for regenerating ORP platinum
electrodes, consisting of 10 g activation
powder and 30 g chlorine powder

SORT/RH

109 730Y

Zobell ORP buffer solution (125 ml)

3682