YSI TruLine ORP 51 User Manual

Page 7

Advertising
background image

7

T 814246

04/2014

TruLine

Commissioning, measuring, testing

Conversion to the
standard hydrogen
electrode potential

U

H

= U

Meas

+ U

Ref

where:

U

H

= ORP voltage, relative to the

standard hydrogen electrode

U

Meas

= measured ORP voltage

U

Ref

= voltage of the reference system relative to the

standard hydrogen electrode

U

Ref

is temperature dependent and can be taken from the following table

(also refer to DIN 38404-6):

T (°C)

U

Ref

[mV]

TruLine ORP 51

T (°C)

U

Ref

[mV]

TruLine ORP 51

0

+221

35

+187

5

+216

40

+181

10

+212

45

+176

15

+207

50

+171

20

+202

55

+165

25

+197

60

+160

30

+192