Commissioning, measuring, testing

TruLine

6

T 814246

04/2014

Commissioning, measuring, testing

Note
For ORP electrodes with platinum electrode, please observe the instructions

in section A

CTIVATING PLATINUM ELECTRODES

.

Commissioning

Prepare the combination electrode for measuring as follows:

 Remove the watering cap from the electrode tip. Possible salt deposits in

the area of the watering cap do not affect the measuring characteristics
and can easily be removed with deionized water.

Note
Please keep the watering cap. It is required for storing the combination

electrode. Always keep the watering cap clean.

 Connect the combination electrode to the meter.

 Measure with the combination electrode according to the operating man-

ual of the meter and observe the following rules while doing so:

General rules for

measuring

 TruLine ORP 51: Make sure the refilling opening for the reference electro-

lyte solution is open.

 Avoid displacement of the sample solution from one measurement to the

next by taking the following measures:

– Briefly rinse the calibration beakers with the solution the beakers are to

be f illed with next.

– Between measurements, rinse the combination electrode with the so-

lution specified below. Alternatively, you can also rinse the combination
electrode with deionized water and then carefully dab it dry.

 Immerse the combination electrode in the solution in a vertical or slightly

tilted position.

 Make sure the immersion depth is correct. The junction must be complete-
ly submersed in the solution. The junction is in the area of the bottom end of
the shaft (see arrow).

The level of the reference electrolyte must be at least 2 cm above the level of
the solution.