For standard logic, that is a solid tone in oil and a pulsing tone in water, remove and

replace the battery without holding down the start button.

To reverse the above logic, hold down the red start button while inserting the battery

drawer. The unit will now produce a solid tone in water and a pulsing tone in oil.

The unit will make a brief audible tone when the battery is first installed, until the auto

off feature powers the unit down into the sleep mode after 15 minutes. When the

unit is on, but not in any liquid, the red LED with flicker briefly every few seconds, so

that the operator knows the unit is turned on.

IV. Basic Reel Features

The dipmeters come equipped with a probe holder and well casing bracket/hook.

When taking a reading it is possible to hang the reel from the well casing and use the

frame leg to guide the tape into the well. The frame leg should protect the tape from

the sharp edges of the well casing. In the event the tape will be exposed to the well

casing’s sharp edges, do not use the reel in the above position but order a separate

tape guide suitable for your specific requirements.

The unit has a reel brake at the back of the frame. To lock the reel, tighten the brake

in the clockwise direction.

If equipped, make certain to use the grounding strap located at the back of the reel.

Additional Features-Pump Monitoring Mode

During a pump test or draw down test, the unit can be used to alert the operator that

the pump will run dry soon.

While the unit is in any liquid, hold down the start button for at least 4 seconds, and

the audio signal will stop and the unit will go into pump alarm mode.

Once the probe is removed from the liquid or the liquid level drops below the probe,

the audio signal will resume.