V. operating the unit in the field – Xylem WL550 OIL WATER INTERFACE METERS User Manual

Page 6

Advertising
background image

Global Water

800-876-1172 globalw.com

- 6 -

To turn off this feature or resume normal operation, simply hold down the start button

for at least 4 seconds.

If you hold down the start button for 4 seconds while the probe is not in a liquid, it will

signal that the probe is in the air and the signal will only stop once the probe is

immersed in a liquid.

V. Operating the unit in the field

The oil/water interface meter can be tested in most containers to detect water. It will

even detect distilled water that contains some dissolved solids. It is more sensitive

to water, since the pins are often coated with oil or product that coats the water

sensing pins. The unit will not typically detect product if there is more than low

ambient light conditions or a white container. This is completely normal and the unit

should NOT be recalibrated, since most wells are very deep and have very little

ambient light.

The water level meter is not as sensitive to water and will not usually detect

commercial quality distilled water.

It is possible to recalibrate the unit for very viscous oils, for a severely scratched

prism assembly or for very conductive water, for example brine or salt water

intrusions.

Adjustment of water sensitivity or hydrocarbon sensitivity

DO NOT CHANGE WITHOUT CALLING FOR TECHNICAL SUPPORT FIRST

There are two small potentiometers located on the circuit board. One is labeled

water. Turn clockwise for very pure/low conductivity water. For saltwater turn

counterclockwise. For white PVC casing or a heavily emulsified layer, turn the fluid

potentiometer counterclockwise. For highly viscous oils, turn the fluid potentiometer

clockwise. Before beginning field work, it is worthwhile to test the unit quickly in any

container to verify it will detect water. The unit will sound continuously if the battery

is low. Since the battery is easily changed without any tools, it is recommended that

a spare battery be included in the carry case/field kit.