6.2 Control Panel

Enter Button (right arrow) -

! Press and release - Toggle between

Rate, Total, and Calibrate screens in
the run mode. Select program
screens in the program mode.

! Press and hold 2 seconds - Enter

and exit program mode. (Automatic
exit program mode after 30 seconds
of no inputs).

MICRO-FLO

Page 8

6.4 Run mode display

1

0

S

ML

R

Min

Body size/range
0 = Field calibrate
1 = 30-300 ml/min
2 = 100-1000 ml/min
3 = 200-2000 ml/min
4 = 300-3000 ml/min
5 = 500-5000 ml/min
6 = 700-7000 ml/min

Function indicator
R = Flow rate indicated
T = Flow total indicated

Rate time base indicator
Min = Rate per minute
Hr = Rate per hour
Day = Rate per day

Display Value

Flow units indicator
ML = Milliliters
OZ = Ounces
GAL = Gallons
LIT = Liters

SetP

Alarm indicator
SetP (steady) = active
SetP (flashing) = alarm
none = not programmed

Cal

Field Calibration indicator
Cal (steady) = active
Cal (flashing) = calibrating
none = factory cal. active

Clear/Cal (up arrow) -

! Press and release - Clear total in the run mode. Scroll through and Select

options in the program mode.

6.3 Flow stream requirements

! The Micro-flo flowmeter can measure fluid flow in either direction.
! The meter must be mounted so that the paddle axle is in a horizontal

o

position - up to 10 off the horizontal is acceptable.

! The fluid must be capable of passing infra-red light.
! The fluid must be free of debris. A 150 micron filter is recommended -

especially when using the smallest body size (S1), which has a 0.031”
through hole.