Oxi 3205

Operation

21

ba75797e01

03/2009

4.3

Sensor-independent settings

4.3.1

Setting in the Configuration menu

To open the Configuration menu, press the <ENTER

_

> key in the

measured value display. After completing the settings, switch to the
measured value display with <M>.

4.3.2

Automatic Stability control

The function, automatic Stability control (AutoRead) continually checks
the stability of the measurement signal. The stability has a considerable
impact on the reproducibility of measured values.

The measured parameter flashes on the display

z as soon as the measured value is outside the stability range

z if you switch over between the measured parameters with <M>.

Menu item

Setting

Description

Language

Deutsch
English
(further)

Select the menu language

Illumination

Auto
On
Off

Switching the display
illumination on/off

Contrast

0 ... 100 %

Changing the display
contrast

Switchoff time

10 min ... 24 h

Adjust the switch-off time

Service information

Hardware version and
software version of the
meter are displayed.

Reset

-

Resets the system settings
to the default values.
For details, see section
4.5.2