6 what to do if, What to do if, Chapter 6 w – Xylem Oxi 3205 Dissolved oxygen (D.O.) meter User Manual

Page 33: 6what to do if

Advertising
background image

Oxi 3205

What to do if...

33

ba75797e01

03/2009

6

What to do if...

Error message,

OFL

Error message,

Error

Sensor symbol flashes

Display

Meter does not react to

keystroke

You want to know which

software

version is in the meter

Cause

Remedy

– Measured value outside the

measuring range

– Use a suitable D.O. sensor

Cause

Remedy

– D.O. sensor contaminated

– Clean D.O. sensor and

replace it if necessary

Cause

Remedy

– Cleaning interval expired

– Recalibrate the measuring

system

Cause

Remedy

– Batteries almost empty

– Replace the batteries (see

section 5.1 M

AINTENANCE

)

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the <ENTER> and
<On/Off> key
simultaneously

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

With <ENTER

_

> and

<

▲><▼> open the menu

Service Information. The
instrument data is displayed.

Advertising