Xylem Oxi 3205 Dissolved oxygen (D.O.) meter User Manual
Page 26

Operation
Oxi 3205
26
ba75797e01
03/2009
Calibration evaluation
After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the
display and in the calibration record.
Calibration in
water vapor
saturated air
(air calibration vessel)
Proceed as follows to calibrate the instrument:
Note
The sponge in the air calibration vessel must be moist (not wet). Leave
the sensor in the air calibration vessel for a time long enough to adjust.
Display
Calibration record
Relative slope
+++
S = 0.8 ... 1.25
++
S = 0.7 ... 0.8
+
S = 0.6 ... 0.7
Error
Error
S < 0.6 or S > 1.25
Eliminate the error according to
chapter 6 W
1
Connect the D.O. sensor to the meter.
2
Put the D.O. sensor into the air calibration vessel.
3
Start the calibration with <CAL>.
The last calibration data (relative slope) is displayed.