Calibration evaluation

After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the
display and in the calibration record.

Calibration in

water vapor

saturated air

(air calibration vessel)

Proceed as follows to calibrate the instrument:

Note
The sponge in the air calibration vessel must be moist (not wet). Leave
the sensor in the air calibration vessel for a time long enough to adjust.

Display

Calibration record

Relative slope

+++

S = 0.8 ... 1.25

++

S = 0.7 ... 0.8

+

S = 0.6 ... 0.7

Error

Error

S < 0.6 or S > 1.25

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

1

Connect the D.O. sensor to the meter.

2

Put the D.O. sensor into the air calibration vessel.

3

Start the calibration with <CAL>.
The last calibration data (relative slope) is displayed.