Xylem Oxi 3210 Dissolved oxygen (D.O.) meter User Manual

Page 36

Advertising
background image

Operation

Oxi 3210

36

ba75795e02

03/2009

4.5.2

Displaying and editing the measurement data storage

The contents of the manual measurement data storage can be shown
on the display.

The measurement data storage has a function to erase the entire
contents.

Editing the data storage

The storage is edited in the menu, Storage & config/ Data storage. To
open the Storage & config menu, press the <F1

_

>[Menu] key in the

measured value display. Open the manual storage directly with the
<RCL> key.

Settings

Display presentation

of a dataset

Quitting the display

To quit the display of stored measurement datasets, you have the
following options:

Menu item

Setting/
function

Description

Data storage /
Manual data storage /
Display

-

Displays all measurement
datasets page by page.

Further options:
z Scroll through the

datasets with <

▲><▼>.

z Quit the display with

<F1>/[Back].

Data storage /
Manual data storage /
Erase

-

Erases the entire manual
measurement data storage.

Note:
All calibration data remains
stored when this action is
performed.

Manual data storage (200)

Data record 3 of 64

30.10.2008 11:24:16

ID number:1

O2 = 7.92 mg/l

25.1 °C AR +++

Sal = 6.5

22.09.2008

08:00

Back