3.4 A

DJUSTMENT

3.4.1 I

NTRODUCTION

To achieve effective compressions, it is very
important that the suction cup is positioned

correctly on the patient (Figure

o

).

The lower edge of the suction cup should be
positioned immediately above the end of the

sternum. The suction cup should be centred over

the sternum (Figure

a

).

3.4.2 A

DJUSTMENT

PROCEDURE

WARNING – MINIMUM PATIENT SIZE

If the suction cup cannot be completely compressed
and lowered enough to make the pressure pad touch

the patient's chest, the patient is too small for

LUCAS to be used.

Continue manual compressions.

To adjust the position of LUCAS:
1.

Set the

ON/OFF

knob to

Adjust

.

2. Lower the suction cup with the height-

adjustment handles until the pressure pad

inside the suction cup touches the patient’s
chest without compressing the chest

(Figure

s

).

If there is a distance between the pressure pad

inside the suction cup and the sternum, LUCAS

cannot be used on the patient(Figure

d

).

Note: By simultaneously holding up the outer

rim of the suction cup you will more easily see or

feel when the pressure pad touches the chest.

3. If the suction cup is not correctly positioned in

relation to the patient, adjust the position of

LUCAS by pulling on the support legs

(Figure

f

). The person assembling the device

determines whether the position is correct.

o

a

>0 mm

s

d

f

Advertising