6 defibrillation, 7 transporting the patient, 1 securing the patient’s arms – Physio-Control LUCAS User Manual

Page 19: Efibrillation, Ransporting, Patient

Advertising
background image

100057-00 F, ©J

OLIFE

AB 2007

19

LUCAS

TM

C

HEST

C

OMPRESSION

S

YSTEM

, I

NSTRUCTIONS

FOR

U

SE

3.6 D

EFIBRILLATION

WARNING – INCORRECT ECG ANALYSIS

Compressions interfere with ECG analysis.

Switch off LUCAS before ECG analysis. Make the

interruption as short as possible

WARNING – INADEQUATE COMPRESSION

Patient’s movements may change adjustment of

suction cup.

Verify position of suction cup after defibrillation,

re-adjust if necessary.

Note: Self-adhesive electrodes should be used during

defibrillation as these make it easier to work with

LUCAS. Position the defibrillator electrodes and

wires so they are not under the suction cup; if

electrodes are already on the patient, you may need

to apply new electrodes.

Defibrillation can be performed when LUCAS is

applied to the patient (Figure

h

).

1.

Apply electrodes and perform the defibrillation
according to manufacturer’s instruction for the

defibrillator.

2. After defibrillation or analysis of the outcome

verify the position of the suction cup and re-

adjust if necessary.

3.7 T

RANSPORTING

THE

PATIENT

3.7.1

S

ECURING

THE

P

ATIENT

S

ARMS

The patient’s arms can be fixed to the device during

transport, using the straps on the support legs

(Figure

j

). This makes it easier to move the

patient, especially if not in a hospital environment.

WARNING – BLOCKED INTRAVENOUS

CATHETER

Do not tighten the patient strap/straps if drugs are

administered to the patient via an intravenous

catheter.

Caution – Do not lift by the straps

The straps are only intended for securing the

patient

s arms to LUCAS.

Apply the straps tightly enough to secure the arms,

but do not overtighten as this may reduce blood

circulation to the patient’s hands.

h

j

Advertising