34

100057-00 F, ©J

OLIFE

AB 2007

LUCAS

TM

C

HEST

C

OMPRESSION

S

YSTEM

, I

NSTRUCTIONS

FOR

U

SE

—A

PPENDIX

B, LUCAS S

TABILIZATION

S

TRAP

1.

Lift the patient’s head carefully and place the

support cushion under the patient’s neck, as
close to the patient’s shoulders as possible.

2. Move the buckles on the device straps towards

the patient’s head.

3. Connect the buckles on the support cushion

straps to the device straps. Make sure that the

straps are not twisted.

Figure 6. Attaching LUCAS Stabilization Strap to
LUCAS
.

WARNING – IMPAIRED VENTILATION OF

PATIENT

Never tighten the support cushion straps so much

that this impairs the ventilation of the patient.

4. Tighten the support cushion straps firmly,

while holding the support legs of LUCAS.

Make

sure that the position of LUCAS External

Cardiac Compressor and the position of the

suction cup on the patient, do not change.

5. Verify that the support cushion straps are

properly fastened to the support legs of LUCAS.

6. Visually check the position of the suction cup:

The lower edge of the suction cup should be

positioned immediately above the end of the
sternum. The suction cup should be centered

over the sternum. Refer to Section 4.2.1 if the

suction cup needs to be adjusted

q

w

-e

r

-t

y

WARNING – LIFTING THE PATIENT

LUCAS Stabilization Strap must never replace a

person supporting the patient’s head. Always lift a

patient and LUCAS according to LUCAS Chest

Compression System, Instructions for Use.

WARNING – TRIPPING ON STRAPS

Mind your step to avoid tripping on the support

cushion straps, or unintentionally tightening the

straps, when lifting the patient.

4.2.1 I

F

THE

POSITION

OF

THE

SUCTION

CUP

IS

NOT

CORRECT

Refer to LUCAS Chest Compression System,

Instructions for Use, for details on adjusting the

position of LUCAS relative to the patient.

1.

Turn the ON/OFF knob to the

(

Adjust

)

position.

2. Release the support cushion straps from

LUCAS.

3. Adjust to correct position of the suction cup on

the sternum.

4. Start compressions.
5. Extend the support cushion straps by pulling

the buckles to the end positions of the straps.

6. Re-attach LUCAS Stabilization Strap, see

Section 4.2.

Figure 7. Adjusting the position of the suction
cup.

q

-w

e

-r