4.1.3 D

ISINFECTION

WARNING – HAZARDS DURING DISINFECTION

Always wear appropriate protective clothing during

disinfection of LUCAS. Follow the handling
instructions from the manufacturer of the

disinfectant.

The normal cleaning procedure is sufficient after

“normal” soiling. If LUCAS is blood-stained or if an
infectious patient has been treated, the device

should also be disinfected using 45% isopropyl

alcohol with added detergent.

4.1.4 D

ISINFECTION

PROCEDURE

After cleaning LUCAS, disinfect it using the

following procedure:

1.

Wipe all outer surfaces of the device using a
cloth wetted with the disinfectant solution.

2. Allow LUCAS to dry before packing into the

bag.

4.2 R

OUTINE

CHECKS

After each use and once a week the following

checks should be performed. Sign each item on the

sheet attached in Appendix A:

1.

Check that the device is clean.

2. Check that the air hose does not have any

cracks and fits tightly to LUCAS.

3. Check that the connector(s) is/are present and

undamaged.

4. Check that the

ON/OFF

knob is in the

Adjust

position.

5. Pull up the release rings so that the claw locks

are open for assembly. Attach the upper part to
the back plate.

6. Check that the compression mechanism can be

adjusted from the uppermost to the lowermost
position without any jarring or sluggishness.

7. Attach the air hose to a air supply and switch

the device on for 1-2 minutes. Check that it is
operating normally. Calculate the number of

compressions per minute. 100 compressions/

min ±10 is acceptable.

4.3 S

TORAGE

Store LUCAS in the carrying bag, in a location
where it is easily accessible to all those who know

how to use the device (see packing instructions in

Section 5). The device should not be exposed to dirt
or moisture during storage.

4.4 S

ERVICE

LUCAS should be serviced once a year to check that

it is functioning properly. Use the original package
when sending LUCAS. For this purpose, store the

package and padding that came with LUCAS when

you first received it.

WARNING – DO NOT OPEN

Do not, under any circumstances, open the casing of

LUCAS, change or modify any external or internal

parts of LUCAS.

Unless otherwise stated, all service and repairs must

only be done by service personnel authorised by

J

OLIFE

AB.

If the above conditions are not adhered to, this may

lead to patient or user injury or death.

Contact your distributor or the manufacturer for

current information on where to send LUCAS for

maintenance (see page 2 for addresses).