CL boards with fixed capacitors are

available as well as the CL board with

the trimmer. Around 20 different CL

boards with fixed C and a target

capacitance trimming function will

make your measurement with load

capacitance easier to handle and

much more reliable with substantially

less error components than other

solutions currently in use.

With the HP E4916A, you can mea-

sure a DUT with load capacitance

using the same test fixtures and load

capacitance as those for the network

analyzer. Thus no correction is

required to correlate your results to

those obtained with network ana-

lyzer.

Unlike other conventional network

analyzers, the HP E4916A searches for

the resonance point (zero-phase

point) in a unique manner for

improved efficiency.

It begins to measure the phase at the

lower end of the search range and

continues its search by changing the

frequency at specific intervals. The

resonance point lies somewhere

between the two adjacent points at

which the phase for one point is

negative and that for the other point

positive. This is the primary search.

Then, the HP E4916A slightly goes

back in the search range and changes

the frequency at smaller intervals

such that the phase becomes closer to

0. The HP E4916A repeats its mea-

surements until the phase is suffi-

ciently close to 0. This is the second-

ary search.

Benefit:

This method provides constant

frequency resolution regardless of the

measurement time, because there are

no needs to specify the number of

measurement points. In addition, the

power level is not interpolated.

Therefore, the specified power is

exactly applied to the DUT at the

resonance point.

Primary Search Range

Secondary Search Range

+

+

θ

θ

Resonance Point Determination by Recursive Search