Specifications – Atec Agilent-E4916A User Manual

Page 6

Advertising
background image

6

Measurement parameters:

Fr, Fs, Fa, FL, CI, Q, C0, C1, L1, R1,

R0, G0, spurious, DLD, EM, filter

bandwidth

(Note that Fr and Fs represent the

frequencies at which the phase is 0

degree and the maximum conduc-

tance is attained, respectively. You

can also set up the HP E4916A to

search for Fr at which |Z| is minimal.)

Output

Frequency range: 1 MHz to 180 MHz/

1 mHz resolution

Frequency accuracy: 2 ppm (can be

improved up to 0.2 ppm through

external reference input)

Inputs (typical)

Measurement range: 1 MHz to

180 MHz/1 mHz resolution

Impedance: 5 to 125

/10 m

resolution

Return loss: 20 dB

Max. input level: 0 dBm

Damage level: 25 Vdc, 20 dBm

Average noise level: -130 dBm/Hz

Cross talk: -100 dB

Dynamic accuracy:

Input level

Level

Phase

accuracy

accuracy

0 to -20 dBm

0.5 dB

-20 to -30 dBm

0.4 dB

-30 to -40 dBm

0.1 dB

0.6°

-40 to -50 dBm

0.05 dB

0.3°

-50 to -60 dBm

0.05 dB

0.3°

-60 to -70 dBm

0.1 dB

0.6°

-70 to -80 dBm

0.5 dB

-80 to -90 dBm

1.0 dB

10°

General Specifications

Power supply: 90 to 132 V or 198 to

264V, 47 to 63 Hz, 100 VA max.

Operating temperature/humidity:

0 to 55

°

C/15 to 95%

Outer dimensions: 320W

×

100H

×

450D mm

Weight: 5.3 kg

Specifications

Others

Interface: HP-IB interface, handler

interface

External reference input: 1/2/5/10

MHz,

±

10 ppm

Equipment supplied with the HP

E4916A:

2 BNC cables (120 cm), 1 operation

manual, 1 user's guide, 4 sample

program disks (VEE run-only

versions included), 1 power cable

Measurement Accuracy

Resonator testing:

Fr: 2 ppm (SPC: Supplemental

Performance Characteristics)

CI: 5% (SPC)

LCR measurements:

Basic accuracy: 3% (SPC)

Impedance measurement range (SPC)

Measurement time (SPC):

The duration of time between the

triggering of the HP E4916A and the

completion of the measurement was

measured. (The time required to

transfer data is not included. The

DLD measurement was made at 5

points; 10 nW, 100 nW, 1

µ

W, 10

µ

W,

and 100

µ

W.)

Frequency

Q

Fr,CI

Fr,CI+

DLD

MEAS

EQV

TIME

100 MHz

83,000 240 ms

1.02 s 890 ms

SHORT

60 MHz

110,000 190 ms 730 ms 775 ms

SHORT

24 MHz

135,000 146 ms 680 ms 705 ms

SHORT

2 MHz

40,000 280 ms

1.64 s

- -

MEDIUM

2 MHz

- -

- -

- -

2.57 s LONG

Note: Measurement time may vary

depending on the resonance

frequency even if two resona-

tors of the same type are

tested.

Accessories and Other

Equipment

HP E4915A Crystal Impedance

Meter

Inexpensive model (no variable

power function available with the HP

E4916A)

Frequency range: 1 MHz to 180 MHz

Output power: -5dBm fixed

Comes with all HP E4916A measure-

ment capabilities other than DLD,

EM, and LCR measurement functions.

HP 41900A

πππππ

Network Test

Fixture

Meets the

IEC444 and JIS

standards. Ideal

for final and

acceptance

inspections.

Can test

resonators with load capacitance

when used in combination with

option 001.

HP41901A SMD

πππππ

Network Test

Fixture

SMD type

π

network test

fixture that

offers superb

repeatability.

Ideal for final

and acceptance

inspections.

Supports 6 types of QIAJ (Quartz

Crystal Industry Association of

Japan) SMDs.